Cookie Notice

This site may use cookies. You have been deemed to have accepted the cookies with continue use this website. Click here for read more.

Philips BV Pulsera (2007)

Philips BV Pulsera (2007)


Multi-application mobile X-Ray with rotating anode tube and 12 inch II

  • Manufacturer

    Philips

  • Model

    BV Pulsera

Philips BV Pulsera (2007)

  • 12 inch Image Intensifier
  • Rotating anode tube
  • 2 x TFT Monitors
  • SW: Options:
    • Electrophysiology
    • Pacemaker
    • Cardiac
    • Vascular Cerebral
    • Vascular HQ
    • Vascular CO2
    • Vascular
    • Abdominal
    • Thorax
    • HQ Orthopaedics
    • Head/Spine
    • Orthopaedics
  • Remote control

Generator

  • 80 kHz High frequency.
  • 7.5 kW full-wave.
  • Up to 120 kVp.
  • Up to 60 mA for radiographic film exposures.
  • Continuous fluoro boost mode up to 20 mA.
  • Pulsed fluoro mode.
  • Pulsed fluorography mode up to 60 mA.
  • Full power from outlet:
    • 50 Hz or 60 Hz.
    • 100V-20A to 240V-10A.

X-Ray Tube

  • Rotating anode X-ray tube.
  • Focal spots: 0.3 mm and 0.6 mm.
  • Anode heat capacity: 300,000 HU.
  • Anode cooling rate: 70,000 HU/min.
  • Housing heat capacity: 1,900,000 HU.
  • Anode diameter: 2.8 in.
  • Anode angle: 10°.

Collimator

  • Iris collimation with dual opposing lead shutters allowing elliptical, round and rectangular collimation.
  • Minimum beam size < 5 X 5 cm2.
  • Iris and shutters are continuously variable.
  • All functions remotely controlled from the C-Arm control panel.

Fluoroscopy Mode

  • Focal spot: 0.3 mm and 0.6 mm.
  • kVp range: 40 – 120 kVp.
  • mA range: 0.2 – 20 mA continuous High Definition fluoro boost mode.
  • Auto and manual modes.
  • Continuous or Pulsed operation.
  • APF varies mA, kVp and camera gain.
  • User specific APF tables.

Pulsed Fluoroscopy Mode

  • Half Dose Fluoroscopy.
  • Quarter Dose Fluoroscopy.
  • Focal spot: 0.6 mm.
  • kVp range: 40 – 120 kVp.
  • mA range: 0.4 – 12 mA.
  • Pulse rate: 6.25 or 12.5 pulses per second.
  • Pulse width: 11 to 37 milliseconds.
  • Computer controlled APF, mA, kVp and camera gain.

Pulsed Fluorography Mode

  • Focal spot: 0.6 mm.
  • kVp range: 40 – 110 kVp.
  • mA range: Up to 60 mA.
  • Pulse rate: 3 to 30 pulses per second. (optional)
  • Pulse width: 8.0 / 9.5 / 11.1 milliseconds.
  • Computer controlled camera iris, mA, kVp, & camera gain.

Digital Exposure Mode

  • Focal spot: 0.6 mm.
  • kVp range: 40 – 110 kVp.
  • mA range: 1.0 – 75 mA.

Radiographic Mode

  • Focal spot: 0.6 mm.
  • kVp range: 40 – 120 kVp.
  • Fixed 60 mA.
  • 0.6 mm - mAs range: 3.2 – 125 mAs.

Video Imaging System

Image Intensifier

  • 12 in Image Intensifier (optional)
    • 12 in.: 1.2 lp/mm
    • 9 in.: 1.4 lp/mm
    • 7 in.: 2.0 lp/mm
    • Tri-mode 12 in/9 in/7 in image intensifier
    • DQE: 65% (typical)

Video Camera

  • High resolution CCD camera.
  • 525 line/frame.
  • Full frame capture.
  • 360° motorized rotation.
  • Left-right image reversal.
  • Top-bottom image reversal.
  • Negative mode.
  • Computer controlled features: (optional)
    • Gain.
    • Blanking.
    • Camera iris.

Video Monitors

  • Dual 17 in high resolution, anti-glare, anti-static.
  • Ambient room light sensor.
  • Multi-LED X-ray “ON” indicator.

Hardcopy Options

  • Thermal printer. (optional)
  • DICOM. (optional)
  • Image Capture Device. (optional)
  • Laser aiming device. (optional)

C-Arm Dimensions

  • Height: 71 / 91 in. (1805 / 2305 mm)
  • Width: 35 in. (880 mm)
  • Length: 76 in. (1930 mm)
  • Weight: 695 lbs 315 kg.

Workstation Dimensions

  • Height: 69 in (1760 mm)
  • Width: 32 in (800 mm)
  • Depth: 37 in (925 mm)