Cookie Notice

This site may use cookies. You have been deemed to have accepted the cookies with continue use this website. Click here for read more.

DRÄGER Fabius Tiro

DRÄGER Fabius Tiro


The Fabius Tiro by Dräger provides everything you need for high quality anaesthesia care in small spaces.

  • Manufacturer

    Dräger

  • Model

    Fabius Tiro

  • System ID

    10000141

DRÄGER FABIUS TIRO ANESTHESIA UNIT

  • Yom: 2011
  • In excellent technical and physical condition
  • Including Dräger Infinity Delta Monitor and Scio Modul
  • SW: 3.37
  • Hours Run Since Last Service: 1046; Total Hours Run: 97383; Total Ventilator Hours Run: 1503

BASE UNIT

Dimensions (W x H x D)

Trolley Version (Cart) with COSY Approx. 30.4 x 53.6 x 33.0 in (77.2 x 136.1 x 83.8 cm)

Wall / Ceiling Version with COSY Approx. 28.4 x 21.9 x 30.5 in (72.1 x 55.6 x 77.5 cm)

Weight and load

Fabius® Tiro Trolley (with COSY) without supplementary cylinders

and vaporizers

256.0 lb (116.0 kg)

Wall and Ceiling Mount (with mounting bracket) 107.4 lb (48.7 kg)

Power and battery backup

Power Input 100 to 240 VAC, 50/60 Hz, 70 VA

Operation time with fully charged batteries 45 minutes minimum (up to 120 minutes)

ANAESTHESIA GAS SUPPLY MODULE

Range of fresh gas flow indicators 0.0 to 12.0 L/min

Total fresh gas flowmeter 0 to 10 L/min. calibrated with 50% O2/ 50% N2O gas mixture

O2 flush max. 75 L/min at 87 psi (6 kPa x 100):

Vapor mount Dräger Interlock* or Selectatec®

optional additional Dräger Interlock* or Selectatec Vapor standby

holder.

* Hosts Auto Exclusion and Interlock Vapor.

VENTILATOR OPERATING SPECIFICATIONS

Ventilator E-vent® Electronically controlled, electrically driven

Operating modes Volume Controlled Ventilation

Pressure Controlled Ventilation (Optional)

Pressure Support (Optional)

SIMV/PS (Optional)

Manual Ventilation

Spontaneous Breathing

Control input ranges

Breathing Frequency (rate) 4 to 60 bpm (1/min)

Positive End Expiratory Pressure (PEEP) 0 to 20 cmH2O (hPa)

Inspiration/expiration ratio (Ti:Te) 4:1 to 1:4

Pressure limiting (Pmax) 15 to 70 cmH2O (hPa)

Tidal Volume (Vt) 20 to 1400 mL in Volume Control

20 to 1100 mL in SIMV/PS

Inspiration pause (Tip:Ti) 0 to 50 %

SIMV Inspiratory time 0.3 - 4.0 sec

Inspiratory pressure (Pinsp) PEEP + 5 to 65 cm H2O (hPa)

Inspiratory Flow (InspFlow) 10 to 75 L/min in Volume and Pressure Control modes

10 to 85 L/min in Pressure Support and SIMV/PS modes

Pressure Support Level (PPS) PEEP + 3 to 20 cmH2O (hPa)

Min. frequency for apnoe-ventilation (Freq. Min.) 3 to 20 bpm and "OFF”

Trigger level 2 to 15 L/min

Integrated Safety Functions Sensitive Oxygen Ratio Controller (S-ORC) guarantees a minimum

O2 concentration of 23% in an O2 /N2O mixture.

N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than

0.2 L/min. Audible and visual (flashing red LED) indication in case

Technical Data

Dräger Fabius® Tiro | 07

O2 pressure drops below 1.38 bar (20 psi) ± 0.27 bar (4 psi).

In case of electricity and battery failure, manual ventilation, gas

delivery and agent delivery are possible. High Pressure Safety

Relief Valve opens at 75 ± 5 cmH2O (hPa). Negative Pressure

Safety Relief Valve opens at -7.5 to -9 cmH2O (hPa).

Ventilator monitoring

Monitoring Continuous monitoring of inspiratory O2 concentration (optional),

breathing frequency, tidal volume expiratory, expiration, minute

volume, peak airway pressure, PEEP, and selection of mean or

plateau pressure. In addition, all fresh gas flow information is

displayed as virtual flow tubes.

Expiratory Minute Volume range 0 to 99 L/min

Control Screen 6.5 in (16.5 cm); black/amber (color mode option)

BREATHING SYSTEM

Optional breathing system heater yes

Volume of entire compact breathing system, with absorber, excl.

hoses, excl. bag

1.7 L

Volume of CO2 absorber Standard: 1.5 L reusable canister, Option: 1.2 L prefilled

Drägersorb CLIC absorber

Mounting on swivel arm, left or right, hight adjustable

GAS SUPPLY AND CONNECTION

Gas Supply O2, N2O, Air / O2, Air

Cylinder Yokes (optional) O2/, O2, Air / O2, N2O - pin-index hanger yokes