This site may use cookies. You have been deemed to have accepted the cookies with continue use this website. Click here for read more.
Dräger
Fabius Tiro
10000141
DRÄGER FABIUS TIRO ANESTHESIA UNIT
BASE UNIT
Dimensions (W x H x D)
Trolley Version (Cart) with COSY Approx. 30.4 x 53.6 x 33.0 in (77.2 x 136.1 x 83.8 cm)
Wall / Ceiling Version with COSY Approx. 28.4 x 21.9 x 30.5 in (72.1 x 55.6 x 77.5 cm)
Weight and load
Fabius® Tiro Trolley (with COSY) without supplementary cylinders
and vaporizers
256.0 lb (116.0 kg)
Wall and Ceiling Mount (with mounting bracket) 107.4 lb (48.7 kg)
Power and battery backup
Power Input 100 to 240 VAC, 50/60 Hz, 70 VA
Operation time with fully charged batteries 45 minutes minimum (up to 120 minutes)
ANAESTHESIA GAS SUPPLY MODULE
Range of fresh gas flow indicators 0.0 to 12.0 L/min
Total fresh gas flowmeter 0 to 10 L/min. calibrated with 50% O2/ 50% N2O gas mixture
O2 flush max. 75 L/min at 87 psi (6 kPa x 100):
Vapor mount Dräger Interlock* or Selectatec®
optional additional Dräger Interlock* or Selectatec Vapor standby
holder.
* Hosts Auto Exclusion and Interlock Vapor.
VENTILATOR OPERATING SPECIFICATIONS
Ventilator E-vent® Electronically controlled, electrically driven
Operating modes Volume Controlled Ventilation
Pressure Controlled Ventilation (Optional)
Pressure Support (Optional)
SIMV/PS (Optional)
Manual Ventilation
Spontaneous Breathing
Control input ranges
Breathing Frequency (rate) 4 to 60 bpm (1/min)
Positive End Expiratory Pressure (PEEP) 0 to 20 cmH2O (hPa)
Inspiration/expiration ratio (Ti:Te) 4:1 to 1:4
Pressure limiting (Pmax) 15 to 70 cmH2O (hPa)
Tidal Volume (Vt) 20 to 1400 mL in Volume Control
20 to 1100 mL in SIMV/PS
Inspiration pause (Tip:Ti) 0 to 50 %
SIMV Inspiratory time 0.3 - 4.0 sec
Inspiratory pressure (Pinsp) PEEP + 5 to 65 cm H2O (hPa)
Inspiratory Flow (InspFlow) 10 to 75 L/min in Volume and Pressure Control modes
10 to 85 L/min in Pressure Support and SIMV/PS modes
Pressure Support Level (PPS) PEEP + 3 to 20 cmH2O (hPa)
Min. frequency for apnoe-ventilation (Freq. Min.) 3 to 20 bpm and "OFF”
Trigger level 2 to 15 L/min
Integrated Safety Functions Sensitive Oxygen Ratio Controller (S-ORC) guarantees a minimum
O2 concentration of 23% in an O2 /N2O mixture.
N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than
0.2 L/min. Audible and visual (flashing red LED) indication in case
Technical Data
Dräger Fabius® Tiro | 07
O2 pressure drops below 1.38 bar (20 psi) ± 0.27 bar (4 psi).
In case of electricity and battery failure, manual ventilation, gas
delivery and agent delivery are possible. High Pressure Safety
Relief Valve opens at 75 ± 5 cmH2O (hPa). Negative Pressure
Safety Relief Valve opens at -7.5 to -9 cmH2O (hPa).
Ventilator monitoring
Monitoring Continuous monitoring of inspiratory O2 concentration (optional),
breathing frequency, tidal volume expiratory, expiration, minute
volume, peak airway pressure, PEEP, and selection of mean or
plateau pressure. In addition, all fresh gas flow information is
displayed as virtual flow tubes.
Expiratory Minute Volume range 0 to 99 L/min
Control Screen 6.5 in (16.5 cm); black/amber (color mode option)
BREATHING SYSTEM
Optional breathing system heater yes
Volume of entire compact breathing system, with absorber, excl.
hoses, excl. bag
1.7 L
Volume of CO2 absorber Standard: 1.5 L reusable canister, Option: 1.2 L prefilled
Drägersorb CLIC absorber
Mounting on swivel arm, left or right, hight adjustable
GAS SUPPLY AND CONNECTION
Gas Supply O2, N2O, Air / O2, Air
Cylinder Yokes (optional) O2/, O2, Air / O2, N2O - pin-index hanger yokes